AND Phoenix GH-300 Laboratory Scale, 320 x 0.0001 g
AND Weighing Phoenix series laboratory scales feature an internal calibration mass with rear mounted motor drive assembly, antistatic glass and multiple weighing modes. Multiple environmental filters, GLP, LIMS, ISO compliant more ...
AND Weighing Phoenix series laboratory scales feature an internal calibration mass with rear mounted motor drive assembly and antistatic glass and multiple weighing modes. Available in capacities from 120 to 250 grams with resolution to 0.01 milligram.
Resolution to 0.01mg
Percentage and Counting Modes
Multiple Weighing Units - weighs in grams, ounces, troy ounces, pennyweight, carat, momme, grains, tola, tael
Tall 251mm (9.88") Weighing Chamber
Built-in Motorized Calibration Weight
Automatic or Manual Full Digital Calibration
One Touch Environmental Filters
GLP/LIMS/ISO Compliant
0.01mg Semi-micro resolution
Internal calibration standard - with bult-in motion driven calibration masses. Calibrates itself when sensing a change in ambient temperature
Full digital calibration - Simplifies calibration by eliminating the need to continuously load and unload calibration weights to adjust Zero and Span
Liquid crystal display - LCD
RS-232C interface - can be used to connect with PC and printer
GLP Compliance - data can be output to ANDs AD8121B printer or computer, including date, time, balance ID number and calibration weight
User Friendly Operation
Windows Communication Software
Balance ID Number
Time and Date Function
Data memory fuction
Automatic Counting Accuracy Improvement - ACAI means continuous recalculation and update of average piece weight ensures optimum count accuracy even with very small parts.
Auto Power On Function
Bluetooth Interface option
Under hook capability
Capacity/Resolution
GH-120
GH-200
GH-300
GH-202
GH-252
Gram
(g)
120x0.0001
220x0.0001
320x0.0001
220x0.0001 / 51x0.00001
250x0.0001 / 101x0.00001
(mg)
120000x0.1
220000x0.1
320000x0.1
220000x0.1 / 51000x0.01
250000x0.1 / 101000x0.01
Decimal Ounce (oz)
4.23x0.00001
7.76x0.00001
11.29x0.00001
7.76x0.00001 / 1.80x0.000001
8.82x0.00001 / 3.56x0.000001
Troy Ounce (ozt)
3.86x0.00001
7.07x0.00001
10.29x0.00001
7.07x0.00001 / 1.64x0.000001
8.04x0.00001 / 3.25x0.000001
Pennyweight (dwt)
77.2x0.0001
141.5x0.0001
205.8x0.0001
141.5x0.0001 / 32.8x0.00001
160.8x0.0001 / 64.9x0.00001
Carat (ct)
600x0.001
1100x0.001
1600x0.001
1100x0.001 / 255x0.0001
1250x0.001 / 505x0.0001
Momme (mom)
32.0x0.0001
58.7x0.0001
85.3x0.0001
58.7x0.0001 / 13.6x0.00001
66.7x0.0001 / 26.9x0.00001
Grain Unit (GN)
1852x0.002
3395x0.002
4938x0.002
3395x0.002 / 787x0.0002
3858x0.002 / 1559x0.0002
Tola (t)
10.29x0.00001
18.86x0.00001
27.44x0.00001
18.86x0.00001 / 4.37x0.000001
21.43x0.00001 / 8.66x0.000001
Tael (TL)
3.20x0.00001
5.87x0.00001
8.53x0.00001
5.87x0.00001 / 1.36x0.000001
6.67x0.00001 / 2.69x0.000001
Percentage Min Div
0.01%, 0.1% or 1% (auto prompt)
Counting Min Weight
0.1 mg
Linearity
0.1 mg
0.1 mg
0.2 mg
0.1 mg/0.02 mg
0.1 mg/0.03 mg
Repeatability/Std Dev
±0.2 mg
±0.2 mg
±0.3 mg
±0.2 mg/±0.03 mg
±0.2 mg/±0.10 mg
Stabilization Time
Approx. 3.5 sec (0.1 mg)/8 sec (0.01 mg)
Sensitivity Drift (10°C-30°C)
±2ppm/°C (when automatic self calibration is not used)
Display Refresh
5 times per second/10 times per second
Pan Size
Ø 90 mm/3.6 inches
Physical Dimensions
(mm)
217(W) x 442(D) x 316(H)
(in)
8.54(W) x 17.4(D) x 12.24(H)
Breeze Break Dimensions
(mm)
164(W) x 148.5(D) x 251(H)
(in)
6.45(W) x 5.84(D) x 9.88(H)
Operating Temperature
5C to 40C (41 F to 104 F), 85% RH or less (no condensation)
AND AD-1683 Static Eliminator more info Price: $1,040.25 In Stock
The AD-1683 generates bipolar ions continuously by DC corona discharge separately from positive and negative discharge electrodes and directs the ionized air onto the charged body to eliminate static electricity. The generated ionsare well-balanced in polarity and can eliminate static electricity regardless of the polarity of the charged body
Static Electricity
Generally, insulators such as plastics are prone to be charged with static electricity when the ambient relative humidity is below 45% RH. This may affect weighing and cause a measurement error of several milligrams. The AD-1683 can eliminate static electricity very effectively
*Note: Please confirm that the AC adapter type is correct for your local voltage and receptacle type. To prevent electrical shock, insert the POWER cable into a grouding-type outlet and earth ground the grounding cable
How to install AD-1683:
Choose the installation site so that the AD-1683 can be placed beside the electronic balance and a space wide enough to place sample (10cm to 30cm from the front of the AD-1683) can be secured. Place the sample in front of the AD-1683 inside the specified area and perform static elimination. Then, weighthe sample using the electronic balance
External Dimensions
Static Eliminating Method
DC corona discharge (Bipolar)
Elimination Range
Distance:Approz. 10 cm to 30 cm from the electode. Area: 30x30 cm from the electode. (See Figure 2.)
Elimination Performance
When Charged 5kV:
Distance
10cm
20cm
30cm
50cm
Elimination Time
1 sec
4 sec
15 sec
70 sec
Ambient Temperature and Humidity Ozone Concentration
0°C to 40°C, 80% RH or less
Positive discharge pin: 0.07PPM
Negative discharge pin: 0.25PPM
(Measured at a distance of 1cm from each discharge pin.)
Power Supply
AC adapter. Please confirm that the AC adapter type is correct for your local voltage and receptacle type
AND AD-8922A Remote Controller more info Price: $525.00 In Stock
Remote Controller for AND Weighing balances
Displays weighing data Remotely controls a weighing instrument with key operations (no access to the internal settings) Outputs weighing data to another external device such as a PC or printer via the RS-232C connection Optional BCD, comparator, and analog outputs available Can be panel-mounted
AND AD-8920A Universal Remote Display more info Price: $250.00 In Stock
Universal Remote Display for AND Weighing balances
Connects to A&D’s balances and scales using either RS-232C or current loop
Includes an AC adapter and a communication cable that matches your instrument
Recognizes the balance/scale baud setting automatically
Large and bright VFD display