Model |
5351010131 |
5351020131 |
5351200131 |
5351210131 |
5351220131 |

$1,999.00
|

$1,999.00
|

$2,269.00
|

$2,525.00
|

$2,525.00
|
Main Functions |
NiBP, Oral Temp |
NiBP, Rectal Temp |
NiBP, SpO2 Masimo |
NiBP, SpO2 Masimo, Oral Temp |
NiBP, SpO2 Masimo, Rectal Temp |
Display |
7" Touchscreen-Display |
BIA module |
Without BIA Module |
Blood pressure measurement range – deflation |
pSYS: 25 - 280 mmHg pDIA: 10 - 220 mmHg pMAP: 15 - 260 mmHg |
Blood pressure measurement range – inflation |
pSYS: 77 - 200 mmHg pDIA: 45 - 190 mmHg pMAP: 56 - 193 mmHg |
Blood pressure measurement accuracy – deflation (clinical validated) |
Mean deviation systole 0.39 mmHg; Standard deviation systole 2.57 mmHg; Mean deviation diastole 0.43 mmHg; Standard deviation diastole 1.73 mmHg |
Blood pressure measurement accuracy – inflation (clinical validated) |
Mean deviation systole 0.94 mmHg; Standard deviation systole 3.84 mmHg; Mean deviation diastole 0.57 mmHg; Standard deviation diastole 3.17 mmHg |
Blood pressure measurement method |
Oscillometric, deflation or inflation measurement |
Blood pressure measurement certainty |
Excess pressure limit 300 mmHg; automatic pressure release at > 300 mmHg |
Blood pressure measurement time |
Normal: 15 - 20 sec / max. 90 sec (adults) |
Blood pressure pulse rate accuracy |
max. +/- 3 bpm |
Blood pressure pulse rate range – deflation |
30 - 240 bpm |
Blood pressure pulse rate range – inflation |
45 - 200 bpm |
Net weight |
5.9 lbs / 2.7 kg |
Product height |
10.3 inch / 262 mm |
Product width |
10.0 inch / 254 mm |
Product depth |
10.9 inch / 278 mm |
Ports |
Ethernet USB 2.0 Wi-Fi |
SpO2 module |
Without SpO2 module |
Masimo SpO2 module |
SpO2 measurement range |
0 - 100 % |
SpO2 measurement accuracy |
SpO2 (resting) 70 % - 100 % + 2 Arms; SpO2 (in motion) 70 % - 100 % + 3 Arms |
SpO2 measurement method |
Pulse oximetry |
SpO2 pulse rate measurement range |
Pulse rate: 25 bpm to 240 bpm Perfusion index measuring range 0.02% - 20.0% |
SpO2 pulse rate measurement accuracy |
PR (no motion) 25 - 240 bpm + 3 digits PR (in motion) 25 - 240 bpm + 5 digits |
Power supply |
Built-in power adapter internal rechargeable lithium-ion battery |
Temperature measuring range |
Direct 30 oC - 43 oC (86 oF - 109.4 oF predictiv 35.5 oC - 42 oC (95.9 oF - 107.6 oF) |
- |
Direct 30 oC - 43 oC (86 oF - 109.4 oF predictiv 35.5 oC - 42 oC (95.9 oF - 107.6 oF) |
Temperature measurement response time |
Predictive (oral) 3 - 10 sec |
Predictive (axillary/rectal) 8 - 14 sec. Direct 60 - 120 sec |
- |
Predictive (oral) 3 - 10 sec |
Predictive (axillary/rectal) 8 - 14 sec. Direct 60 - 120 sec |
Temperature measurement accuracy |
Direct + 0.1 oC (+ 0.2 oF) Predictive (for 95 % of measurements) + 0.1 oC (+ 0.2 oF) |
- |
Direct + 0.1 oC (+ 0.2 oF) Predictive (for 95 % of measurements) + 0.1 oC (+ 0.2 oF) |
Temperature measuring method |
Oral/axillary (direct or predictive) |
axillary/rectal (direct or predictive) |
- |
Oral/axillary (direct or predictive) |
axillary/rectal (direct or predictive) |
Temperature module |
Filac™ temperature oral/axillary |
Filac™ temperature axillary/rectal |
- |
Filac™ temperature oral/axillary |
Filac™ temperature axillary/rectal |